Multitron Standard - Operating Manual
Operation
17 March 2021
Page
53
of
114
6.2.2
Removing the Tray
The device is fitted with a tray lock to hold the tray securely in place
on the table. When the door is opened fully, the lock hooks (1) on
the table release the tray. The ejection levers under the table lift the
tray out of the table’s lock cones (2). The tray is now free and can be
pulled out of the incubation chamber by hand.
The tray is removed from the incubation chamber using slide rails (2)
and the cylindrical ball casters inside them (1), which are located on
the inside of the door. Two screw-mounted pins (3) in the slide rails
limit how far the tray can be pulled out. When pulled out all the way,
the tray lies partially on the door and partially on the table.
To remove the tray, proceed as follows:
Procedure
1.
Open the device door all the way.
The safety mechanism unlocks the lock hooks automatically
when the door is opened, and lifts the tray out of the rear
cones.
2.
Lift the tray up by its handle.
3.
Pull the tray over the front stop bar and onto the open door.
INFORMATION
If you cannot pull the tray out, check whether the door is open all
the way. Push the door down as far as it will go.
1
2
2
1
3