Multitron Standard - Operating Manual
Installation
17 March 2021
Page
49
of
114
5.3
Minimum Distances to the Device
When setting up the device, the following minimum distances must
be observed to ensure adequate ventilation and access to the most
important connections:
Minimum distances to the device
Placing the device on a bench
When placing the device on a bench, observe the following points:
To prevent the bench from vibrating, use a heavy bench with
diagonal braces as the minimum (ideally, use a stable labora-
tory bench). Alternatively, the bench can be affixed to the wall.
The minimum required depth for the bench is 815 mm.
You must leave a ventilation gap of at least 100 mm between
the device and rear wall.
If shelves or cupboards are installed above the bench, you must
leave a minimum distance of 100 mm.
No gas outlets, shelves or cupboards must be in the way.