Multitron Standard - Operating Manual
Setup and Function
17 March 2021
Page
27
of
114
INFORMATION
For information on the technical data for the cooling system, see
chapter 11.3.2 "Temperature Parameter", page 106.
3.4
Interior Lighting
The device features interior lighting. The interior lighting is switched
on automatically as soon as a button is pressed or the door is
opened. After 60 seconds without input, the lighting switches off
again immediately.
The interior lighting comprises two halogen lamps, which are located
on the inside of the door, near the top. In case of a defect, the halo-
gen lamps can be replaced by the owner.
3.5
Connections and Interfaces
3.5.1
Mains Connection
The mains connection is located on the left side of the device, to-
wards the rear. Two different versions of the device are available for
different mains voltages:
230 V 50/60 Hz
115 V 60 Hz
Two fuses immediately adjacent to the power connection protect the
device from impermissibly high power input.
The country-specific power cable required for connecting the power
is included in the device’s scope of delivery. If the power cable is de-
fective, replace it with a power cable of the same type.
Prior to connecting the device, make sure that the voltage values of
the device match those of the local power supply. The mains connec-
tion must be easily accessible at all times so that the device can be
disconnected from the power supply quickly in case of an emergency.
For information on the electrical connection values, see chapter
11.2.2 "Electrical Connection and Performance Values", page 102.