6 | Operation
INFICON
54 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
ð
The internal background is created from the remaining gas, which has not yet
been pumped out. Very clean systems show a background in the range of 1 x
10
-11
mbar l/s. Under normal conditions there is a background of 1 x 10
-10
mbar l/s or 1x10
-9
mbar l/s. For the mathematical treatment of the internal
background, see "Change background suppression [
69]”.
4
Save
.