6 | Operation
INFICON
74 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
6.4 Measuring
ü
The inlet flange on the upper side of the device is prepared for the pending
measurement. Your test specimen or the SL200 sniffer line is normally connected
there.
ü
Alternatively, the sniffer line SL3000 is connected to the back of the device
(optional). In this case, the inlet flange on the top of the device should be closed
with a blank flange.
ü
Possible alternatives to the operating possibilities on the device are set up
(optional):
I/O module, bus module, remote control, wireless LAN connection to the notebook
or tablet. Also see "Accessories and spare parts [
112] " and "Operate leak
detector via web browser (wireless LAN) [
1
Switch on the leak detector via the power supply switch.
2
Make sure that the desired operation mode is set (see "Select operation mode
61]").
3
Make sure that the correct basic settings and the settings for the current
measurement are carried out. See "Basic settings [
4
When performing a calibration, note the 20 minute warm-up time. See
5
In standby mode, initiate the measurement process by pressing the START
button.
ð
If you want to ensure that a current internal background is used in the
following measuring operation, please observe the notes in chapter “Change
69]“.
6
Keep track of the measurement results either as line, bar or circle graph, see
"Presentation of the measurement screen [
25]".
7
To display the maximum leak rate (Q
max
) on the measurement screen during a
measurement, press the START button again. Also see "START button [
and "Assembly of the touchscreen [
8
To measure small leak rates more easily, press the ZERO button when
required, see "Set and use the function ZERO [
67]".
9
Record the measured values if necessary, see "Switch on data recording and
75]" and "Assembly of the touchscreen [
25]".
10
Switch the instrument off.
See also
2
Use HYDRO•S (UL3000 Fab PLUS) [
2
Содержание ULTRATEST UL3000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 111: ...INFICON Decommissioning the measuring instrument 9 UL3000 Fab Operating instructions iina73en1 08 1910 111 140 ...
Страница 132: ...11 Appendix INFICON 132 140 UL3000 Fab Operating instructions iina73en1 08 1910 11 2 CE Declaration of Conformity ...
Страница 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Страница 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 137: ...INFICON Index UL3000 Fab Operating instructions iina73en1 08 1910 137 140 Z ZERO 74 Button 23 Factory setting 67 ...
Страница 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 139: ......
Страница 140: ......