10 | Accessories and interfaces
INFICON
118 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
Pressure p1 / Pressure
p2
1 ... 10 V; 0.5 V / decade;
logarithmic
1 V = 1 x 10
-3
mbar
Leak rate mantissa
1 ... 10 V; linear; in the
selected unit
Useful only if the other analog output is assigned
"Leak rate exponent".
Leak rate exponent
1 ... 10 V; 0.5 V / decade;
Step function;
1 V = 1 x 10
-12
; in selected
unit
Useful only if the other analog output "Leak rate
mantissa" or "Leak rate mantissa hyst." is assigned.
Linear leak rate
0 ... 10 V; linear;
in the selected unit
The upper limit (= 10 V) is set via the parameter
"Exponent upper limit” in whole decades. The lower
value is always 0 (leak rate), which corresponds to 0
V output voltage.
This setting is for both analog outputs, if an
appropriate output function is selected. Depending on
the selected leak rate unit there is a different absolute
limit.
The selected range can be additionally narrowed by
the limits, which is valid for all interfaces.
Leak rate log.
0 ... 10 V; logarithmic;
in the selected unit
The upper limit (= 10 V) and the scale (V / decades)
are set via the parameters "Exponent upper limit" and
"Analog output scaling".
For example:
Upper limit set to 1 x 10
-5
mbar l/s (= 10 V). Scale set
to 5 V / decade. The lower limit is thus 1 x 10
-7
mbar l/
s (= 0 V). The logarithmic output function of both the
slope in V / decade as well as the upper limit (10 V
limit) can be set. This results in the minimum
displayable value. The following slopes are available:
0.5, 1, 2, 2.5, 3, 5, 10 V/The higher the selected slope
value, the smaller the displayable area. The upper
limit is the same for both analog outputs. Depending
on the selected leak rate unit there is a different
absolute limit.
Set by interface
The output voltage can be specified for tests with the LD log command 221.
Leak rate mantissa hyst.
0.7 ... 10 V; linear;
in the selected unit
Useful only if the other analog output is assigned
"Leak rate exponent". Through an overlap of the
mantissa in the range 0.7 to 1.0, a constant jumping
between two decades is prevented. 0.7 V
corresponds to a leak rate of 0.7 x 10
-x
. 9.9 V
corresponds to a leak rate of 9.9 x 10
-x
.
Содержание ULTRATEST UL3000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 111: ...INFICON Decommissioning the measuring instrument 9 UL3000 Fab Operating instructions iina73en1 08 1910 111 140 ...
Страница 132: ...11 Appendix INFICON 132 140 UL3000 Fab Operating instructions iina73en1 08 1910 11 2 CE Declaration of Conformity ...
Страница 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Страница 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 137: ...INFICON Index UL3000 Fab Operating instructions iina73en1 08 1910 137 140 Z ZERO 74 Button 23 Factory setting 67 ...
Страница 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 139: ......
Страница 140: ......