6 | Operation
INFICON
60 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
If the option "calibration request" is active, you will get a calibration request
subsequently in the following cases:
• The period of the device is longer than 30 minutes and the temperature deviation
compared to the last calibration is larger than 5 Kelvin.
• There was no calibration within the last 24 hours in the currently selected
operation mode.
ü
Operator or Supervisor rights
1
> Setup > Measurement > Miscellaneous
2
To switch on further calibration requests activate the option “Calibration
request“. To switch off deactivate this option.
3
Save
.
6.2.14 Extending or reducing operation options
You can always operate the device via touchscreen. Furthermore, you can change the
factory setting, which enables the operation via buttons on the control panel and via
the remote control.
ü
Operator or Supervisor rights
1
> Setup > Measurement > Miscellaneous
2
Under “Local control“ you make your choice in the selection field between:
ð
Released.
The buttons START, STOP and ZERO on the control panel and also the
buttons of the remote control are released. The operation via touchscreen
can be disabled, however, if no user is logged on.
ð
Release after login.
The buttons START, STOP and ZERO on the control panel and also the
buttons of the remote control can be operated after a user has logged in to
the device with a valid PIN.
ð
Locked.
The device can only be operated via the touchscreen. The START, STOP
and ZERO buttons on the control panel are locked. The buttons of the remote
control can also not be used.
3
Save
.
6.2.15 Use Favorites
To shorten the time spent searching menus for frequently used functions, create freely
assignable menu keys in the "Favorites" window.
Содержание ULTRATEST UL3000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 111: ...INFICON Decommissioning the measuring instrument 9 UL3000 Fab Operating instructions iina73en1 08 1910 111 140 ...
Страница 132: ...11 Appendix INFICON 132 140 UL3000 Fab Operating instructions iina73en1 08 1910 11 2 CE Declaration of Conformity ...
Страница 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Страница 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 137: ...INFICON Index UL3000 Fab Operating instructions iina73en1 08 1910 137 140 Z ZERO 74 Button 23 Factory setting 67 ...
Страница 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 139: ......
Страница 140: ......