Operation | 6
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
85 / 140
V10
V9
V8
V7
V5
V4c
V4b
V4a
V3
V2b
V2a
V1
p4
p2
p1
MS
TL
1
2
3
4
5
7
8
6
9
10
Fig. 13: The most important components of the vacuum diagram in UL3000 Fab devices
1 Mass spectrometer
6 Sniffer line
2 Internal calibration leak
7 Pressure measuring point for flow
monitoring
3 Inlet flange
8 Fore pump
4 Inlet pressure gauge
9 Fore-vacuum pressure gauge
5 V1 … V10: Solenoid valves
for controlling gas flows
10 Turbo molecular pump
6.12 Resetting to factory settings
You can restore the device to factory settings.
Loss of settings and measurement data
After resetting to factory settings only the manufacturer factory settings are in the
memory of the device. The software of the device is not reset when restoring the
factory settings.
ü
Supervisor - rights
1
> Setup > General > Reset
2
Make your choice:
ð
If you want to reset all settings, press the "Reset settings" button.
ð
If you want to delete all recorded measurement data and user accounts,
press the "Delete data" button.
Содержание ULTRATEST UL3000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 111: ...INFICON Decommissioning the measuring instrument 9 UL3000 Fab Operating instructions iina73en1 08 1910 111 140 ...
Страница 132: ...11 Appendix INFICON 132 140 UL3000 Fab Operating instructions iina73en1 08 1910 11 2 CE Declaration of Conformity ...
Страница 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Страница 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 137: ...INFICON Index UL3000 Fab Operating instructions iina73en1 08 1910 137 140 Z ZERO 74 Button 23 Factory setting 67 ...
Страница 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 139: ......
Страница 140: ......