INFICON
Operation | 6
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
69 / 140
2
If necessary activate the option "Automatic purge".
ð
If you switch off the automatic purging, you can switch purging on or off in the
standby window by pressing
.
3
Save
.
Regeneration
It may happen that too much helium has entered the interior of the device and no
accurate measurements are possible (helium contamination). The regeneration
provides the user with an automatic start-stop-vent cycle to degrade such elevated
helium background.
An active regeneration is shown in the display.
Regeneration is automatically deactivated after 60 minutes.
ü
Operator or Supervisor rights
ü
You have not activated the setting "No ventilation" under "Ventilation mode", see
above.
►
> Regeneration
ð
You can start the regeneration in the opened window and stop it as desired or
wait for the automatic deactivation of the regeneration after 60 minutes.
6.3.12 Change leak rate filter
The leak rate filter “I•CAL” is preset. I•CAL ensures that the averaging time is optimally
adapted to the leak rate range.
The abbreviation I•CAL stands for “Intelligent Calculation Algorithm for leak rates”.
This ensures that the signals are averaged at optimal time intervals and this is based
on each leak rate range. I•CAL also determines interference spikes, which bear no
relevance to the leak rate signals and delivers unusually short reaction times even
with low leak rate signals. The algorithm used offers excellent sensitivity and response
time.
With the leak rate filter "Fixed" there is also a filter with a fixed time constant available.
Procedure
ü
Supervisor - rights
1
> Setup > Measurement > ZERO and filter
2
Make a selection between “I•CAL“ and “Fixed” under “Leakage rate filter”.
3
Save
.
6.3.13 Change background suppression
The internal background suppression is preset.
The measurement system of the leak detector contains a residual amount of helium
and hydrogen even without connection to a test chamber. This creates an internal
measurement signal component already before pressing the START button.
Содержание ULTRATEST UL3000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 111: ...INFICON Decommissioning the measuring instrument 9 UL3000 Fab Operating instructions iina73en1 08 1910 111 140 ...
Страница 132: ...11 Appendix INFICON 132 140 UL3000 Fab Operating instructions iina73en1 08 1910 11 2 CE Declaration of Conformity ...
Страница 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Страница 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 137: ...INFICON Index UL3000 Fab Operating instructions iina73en1 08 1910 137 140 Z ZERO 74 Button 23 Factory setting 67 ...
Страница 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 139: ......
Страница 140: ......