INFICON
Operation | 6
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
67 / 140
6.3.10 Set and use the function ZERO
Why should I use
ZERO?
To be able to clearly measure small leakages the function ZERO should be used.
With each leak detection there is a "Background signal" (see "Definition of terms
8]") that disrupts the search or the measurement of leakages.
• To hide the background signal, activate the function ZERO.
• You can also use ZERO to hide a currently displayed leakage, which interferes
with searches for other or even smaller leaks.
A currently displayed leak is hidden through ZERO.
By using the function ZERO not only is the background signal hidden, but also the
representation of a current leak.
► If you want to prevent this, select the function ZERO only if not measuring a leak at
the same time.
Activate or deactivate
ZERO
ZERO is activated by factory settings.
ü
Supervisor - rights
1
> Setup > Measurement > ZERO and filter
2
If necessary change "ZERO mode". The factory setting is “I•ZERO 2.0”.
ð
If you select "OFF”, the ZERO button does not work.
3
Save
.
ð
If ZERO is activated, start the function by pressing the ZERO button. See also
23]".
How do I switch the
ZERO function ON?
Briefly press the ZERO button on the control panel or the touchscreen
.
How do I switch the
ZERO function OFF?
Press the ZERO button on the control panel for more than 2 seconds or on the
touchscreen
.
Operation mode
“Vacuum”, effect on
ZERO
After pressing the ZERO button, the displayed background signal is dimmed as a
function of the current drift. Typically, the display limit falls in several steps. The size of
the current display limit is a dimension with which leakages can currently be
measured. When measuring, the current display limit should be at least one decade
below the expected measured value.
Содержание ULTRATEST UL3000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 111: ...INFICON Decommissioning the measuring instrument 9 UL3000 Fab Operating instructions iina73en1 08 1910 111 140 ...
Страница 132: ...11 Appendix INFICON 132 140 UL3000 Fab Operating instructions iina73en1 08 1910 11 2 CE Declaration of Conformity ...
Страница 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Страница 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 137: ...INFICON Index UL3000 Fab Operating instructions iina73en1 08 1910 137 140 Z ZERO 74 Button 23 Factory setting 67 ...
Страница 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Страница 139: ......
Страница 140: ......