7 – 21
Transpector SPS Operating Manual
4
The ion source can now be removed by lifting it off the quadrupole assembly.
Install a new ion source by reversing the above steps. (See
.)
Figure 7-22 Ion Source and quadrupole assembly
5
Using an ohm meter, check that the filament is not shorted to ground or to any
of the ion source plates and that the two filament leads show a filament
resistance of approximately 1
at the feedthrough (approximately 0.5
at the
lead connectors).