8 – 7
Transpector SPS Operating Manual
Filament current error
Insufficient vacuum
Verify pressure is less than
5E-4 Torr at the ion source
Filament broken
Verify integrity of filament
Filament shorted
Verify that filament is not
shorted to ground
Cable box not fully seated
on Transpector electronics
module
Ensure Cable box is fully
seated on Transpector
electronics module
Focus error
Unable to start or maintain
focus voltage
Reset electronics module
Return electronics module
to INFICON for repair
No spectra
Emission is OFF
Turn emission ON
EM is ON, when operating
Turn EM OFF
Contaminated sensor
Degas, or service sensor
Replace sensor
Electronics failure
Return to INFICON for
repair
Pressure too low for FC
Use EM detector
EM voltage too low
Increase voltage
Transpector SPS
electronics module not fully
engaged on sensor
Make sure that the
Transpector SPS
electronics module is
pushed completely on
sensor
Mass calibration
Adjust mass calibration
Table 8-1 Symptom—Cause—Remedy Chart (continued)
SYMPTOM
CAUSE
REMEDY