10 – 8
Transpector SPS Operating Manual
V2
—(not configured on GUI) optional auxiliary pneumatic valve control
V3
—(not configured on GUI) optional auxiliary pneumatic valve control
V4
—(not configured on GUI) optional auxiliary pneumatic valve control
V5 (Calibration)
—calibration valve (available if a calibration gas is
installed)
To open the valve, click on the
Valve Status
indicator in the Transpector
SPS System GUI at the bottom of the screen. The
Valve Status
indicator
will switch from dark blue to green.
8
Click
Start Configuration Monitor
. The sensor will begin collecting data with
the configuration set up using the chosen data acquisition parameters.
9
If the data is satisfactory, log the parameters and save the configuration. Use
these data acquisition parameters in any data collection method.
10
If the data is unsatisfactory, repeat this procedure with different parameters.
CAUTION
CIS Pressure (Torr) is calibrated in the factory at
40 eV electron energy at 200 µA emission current.
Changing electron energy and/or emission current will
cause incorrect CIS pressure readings.
10.6 Transpector SPS Sensor Acquisition Defaults
10.6.1 What to Acquire (RGAs)
The
What to Acquire
parameters decide the data that the Transpector SPS will
collect. This menu is very similar for both Transpector SPS and Open Ion Source
RGAs.
10.6.1.1 Acquisition Modes
There are two different acquisition modes:
Spectrum (Scans)
and
Selected Masses (Bins)
. (See
.)
Figure 10-12 Acquisition modes
Spectrum (Scans)
is the default mode of data acquisition in
FabGuard Explorer. This mode will take data across a user-defined
mass range. (See
.)