IC6 Operating Manual
Chapter 11
Applications
11.1 Multiple Sensor Deposition Control
The conventional method of controlling thin film depositions is to use one crystal
sensor for each material source. In addition to this conventional method, the IC6
can make use of multiple sensors for control of each material source.
To use multiple sensor deposition control, elect the appropriate Sensor Option for
as many sensors as desired (up to 8). For conventional deposition control (one
sensor per material source) select one sensor.
The multiple sensor deposition control feature is designed to monitor a source's
flux distribution using many sensors. Multiple point sensing improves thickness
reproducibility by controlling an aggregate (averaged) deposition rate. Two
important elements of a deposition process contribute to layer-to-layer thickness
variances and cannot be addressed by a single point controller. First, a deposition
source never has the idealized (cosine) flux distribution. Second, the flux
distribution is dynamic, changing as the deposition progresses.
The IC6, when using more than one sensor for each material, calculates an
aggregate rate and uses it for rate control and thickness termination. This
aggregate rate is comprised of rate information from up to 8 sensors, each sensor
having its own weighting and tooling factors.
As the source distribution changes, any individual sensor may detect more (or less)
of the source flux. However, the aggregate rate is held constant. Each substrate,
contained in a rotating planetary substrate holder, is then exposed to the constant
aggregate rate, eliminating thickness variations due to a change in the source's flux
distribution.
This concept is illustrated in
, which also shows the preferred
placement of the sensors relative to the deposition source and substrates.
Содержание IC6
Страница 1: ...O P E R A T I N G M A N U A L IC6 Thin Film Deposition Controller IPN 074 505 P1E Cover Page ...
Страница 2: ......
Страница 8: ......
Страница 20: ...TOC 12 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 42: ...1 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 134: ...6 6 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 148: ...8 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 170: ...9 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 236: ...11 2 IPN 074 505 P1E IC6 Operating Manual Figure 11 1 The essential advantage of multi point deposition sensing ...
Страница 268: ...13 2 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 312: ...15 40 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 338: ...A 10 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 342: ...1 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...