10 - 42
IP
N 07
4-
50
5-
P1
E
IC6 Operating Manual
6
(0x06)
Status Messages
8
<Byte><Byte><Byte>
<Byte><Byte><Byte>
<Byte><Byte>
. One
bit is set for each error.
Multiple bits may be
set.
7
(0x07)
Status Message
Designators
<Status
Message
Word>
8 bytes
1-13
(Can have
12 DAC
recorders)
<Byte>|
<Byte>....<Byte>
Status Message Word
corresponding to one of
the status messages,
as encoded in the SG 6
command.
This indicates which status message designators are to be returned. Note that only one bit may
be set, that is, only the designators for one error can be requested per command.Returns a string
of bytes. First byte is the number of bytes following. The following bytes indicate the designators
for the requested status message.If there are no designators, the first byte returned will be zero.
8
(0x08)
Cause of Stop
1
<Byte>
. If
not currently in Stop,
will return "Data Not
Available" error.
9
(0x09)
Transient
Messages
1
<Byte>
10
(0x0a)
Sensor Status
8
<Byte><Byte><Byte>
<Byte><Byte><Byte>
<Byte><Byte>
One byte for each
sensor Range between
0 and 5.
0 = Good Crystal and active. 1 = Failed Crystal and active. 2 = Invalid Measurement on Crystal
and active. 3 = Good Crystal and inactive. 4 = Failed Crystal and inactive.
5 = Invalid Measurement on Crystal and inactive.
11
(0x0b)
Timer Status
Timer Number
<Integer>
1 - 20
1
<Byte>
Response:0 = Timer
inactive1 = Timer active
12
(0x0c)
Timer Value
Timer Number
<Integer>
1 - 20
4
<Integer>
Response is in tenths of
seconds. Max Time =
99:59:59.0 (3599990
tenths of seconds)
Table 10-12 Status general commands (continued)
SG
Comm
and ID
Description
Option
Number
of bytes
in the
response
Data Returned
Notes
Содержание IC6
Страница 1: ...O P E R A T I N G M A N U A L IC6 Thin Film Deposition Controller IPN 074 505 P1E Cover Page ...
Страница 2: ......
Страница 8: ......
Страница 20: ...TOC 12 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 42: ...1 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 134: ...6 6 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 148: ...8 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 170: ...9 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 236: ...11 2 IPN 074 505 P1E IC6 Operating Manual Figure 11 1 The essential advantage of multi point deposition sensing ...
Страница 268: ...13 2 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 312: ...15 40 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 338: ...A 10 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 342: ...1 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...