5 - 12
IP
N 07
4-
50
5-
P1
E
IC6 Operating Manual
NOTE:
After a crystal fails due to exceeding the Counts value, the
Crystal Fail can be cleared by installing a new crystal or
pressing the F3 key to Clear S & Q counts on the Sensor
Information screen. This action must be repeated for each
sensor used by cursoring to the appropriate sensor number.
STABILITY SINGLE
. . . . . . 0 Hz and 25 to 9999 Hz (Values 1 to 24 are
not allowed)
This sets the maximum permissible increase in frequency from one
measurement to the next that will cause a crystal fail. The default value
is 0 Hz which disables the function.
STABILITY TOTAL
. . . . . . . 0 Hz and 25 to 9999 Hz (Values 1 to 24 are
not allowed)
This sets the maximum total permissible increase in frequency
accumulated during the active layer that will cause a crystal fail. The
default value is 0 Hz which disables the function.
NOTE:
After a crystal fails due to exceeding the allowable Hz value,
the Crystal Fail can be cleared by installing a new crystal or
pressing the F3 key to Clear S & Q counts on the Sensor
Information screen. This action must be repeated for each
sensor used by cursoring to the appropriate sensor number.
5.1.5 Pre/Post Screen Deposit Page Parameters
MATERIAL NUMBER
. . . . . . . . . . . 1 to 32
Change the number to edit or create a different material
NOTE:
The first power ramp consisting of Pre-Cond Rise and Pre-Cond Soak is
skipped if the previous layer had the same source, same crucible and its
Idle power is nonzero.
PRECON SOAK POWER
. . . . . . . . 0.00 to 99.99%
This parameter is usually set to the power level at which the source material
just begins to glow. The IC6 ramps the power level from zero to Pre-Cond Soak
Power linearly over the time period Pre-Cond Rise Time. The default value is 0.
PRECON RISE TIME
. . . . . . . . . . . . 00:00 to 99:59 min:s
This parameter provides the time period over which the source power is
ramped from 0 to Pre-Cond Soak Power. The default values is 00:00.
PRECON SOAK TIME
. . . . . . . . . . . 00:00 to 99:59 min:s
This parameter provides the time period for which the IC6 holds at
Precon Soak Power. The default value is 00:00.
Содержание IC6
Страница 1: ...O P E R A T I N G M A N U A L IC6 Thin Film Deposition Controller IPN 074 505 P1E Cover Page ...
Страница 2: ......
Страница 8: ......
Страница 20: ...TOC 12 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 42: ...1 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 134: ...6 6 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 148: ...8 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 170: ...9 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 236: ...11 2 IPN 074 505 P1E IC6 Operating Manual Figure 11 1 The essential advantage of multi point deposition sensing ...
Страница 268: ...13 2 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 312: ...15 40 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 338: ...A 10 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 342: ...1 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...