7 | Warning and error messages
INFICON
64 / 90
HLD6000-Operating-instructions-kinb43en1-05-(2107)
No.
Message
Possible error sources
Troubleshooting
W110
Clock not set
Clock jumper not set, battery
empty or clock defective.
Contact INFICON customer service.
W111
Many EEPROM write
cycles within the last 6
minutes
• Within the last 6 minutes too
many write commands were
sent to the device from an
external controller
• Too many settings were
changed within the last 6
minutes
• Check the programming of your
external control
• Change settings only if
necessary
W122
No response from bus
module
Connection to BUS module
interrupted
Check the connection to the BUS
module.
W125
I/O module no longer
connected
Connection to I/O module was
interrupted.
Check the connection to the I/O
module.
W126
Protocol not supported
by I/O module software
The I/O module software does not
support the selected HLD5000
protocol.
Make a software update of the I/O
module to a newer version.
W127
Wrong bootloader
version
The boot loader software is not
compatible with the application.
Contact INFICON customer service.
E130
Sniffer line not
connected
The sniffer line cannot be
addressed by the basic unit.
• Check the connection of the
sniffer line with the basic unit
(disconnect and reconnect; if
possible try a different sniffer
line).
• If the problem persists, contact
the INFICON customer service.
E131
Wrong parameter in the
sniffer line
The parameters stored in the
sniffer line are wrong.
Contact INFICON customer service.
E132
Unsupported old sniffer
line
An old sniffer line is connected
which is not supported.
Use a current sniffer line.
E133
EEPROM error in sniffer
line
The EEPROM of the sniffer line is
defective
Contact INFICON customer service.
E134
Protocol error while
communicating with the
sniffer line
The interface to the sniffer line is
not working correctly.
• Check the connection of the
sniffer line with the basic unit
(disconnect and reconnect; if
possible try a different sniffer
line).
• If the problem persists, contact
the INFICON customer service.
Содержание HLD6000
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Страница 84: ...10 Appendix INFICON 84 90 HLD6000 Operating instructions kinb43en1 05 2107 10 3 CE Declaration of Conformity ...
Страница 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Страница 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Страница 89: ......
Страница 90: ......