INFICON
Operation | 6
HLD6000-Operating-instructions-kinb43en1-05-(2107)
53 / 90
2
If you want to test a weld seam or similar, you must guide the tip along the path
at a speed of less than 2.5 cm/s.
3
When you check a spot, hold the sniffer line to it for least 1 second.
4
If you have activated the button function on the sniffer probe, see “Setting up the
46]“, then you have the following options:
ð
If the button is not pressed on the sniffer probe: Measurement takes setpoint
1 into consideration,
ð
If the button is pressed on the sniffer probe: Measurement takes setpoint 2
into consideration.
If you have activated the "Screenshot" function, see "Switching the "Screenshot"
43]", save a screenshot by pressing the button on the sniffer
probe.
If the button function is enabled on the sniffer probe at this time, the screenshot is also
saved when switching to setpoint 2.
If there is a leak, it will be reported in the displays with LEDs in the handle and -
depending on your settings - also accompanied by an acoustic signal.
6.5 Measurement data
6.5.1 Measurement data recording
You can save measurement data as files in TXT format. The device creates a new
TXT file every hour.
There is approx. 16 MB of free internal memory available. This is sufficient when
setting up a 500 ms recording interval for approx. 24 hours.
Alternatively, measured data can be saved on a USB stick up to 32 GB (formatted in
the FAT file system).
Starting recording
1
> Recorder > Recorder settings
ð
Alternatively, when on the touchscreen select the button
.
2
Select from the following settings:
"Memory location": "USB" or "internal"
"Record interval": "100 ms", "200 ms", "500 ms", "1 s", "2 s" or "5s"
3
If "USB" is selected as the memory location then connect a USB stick with the
USB port on the device.
4
Select the "On" button under "Data Record".
5
Start data recording by selecting the button
.
Содержание HLD6000
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Страница 84: ...10 Appendix INFICON 84 90 HLD6000 Operating instructions kinb43en1 05 2107 10 3 CE Declaration of Conformity ...
Страница 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Страница 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Страница 89: ......
Страница 90: ......