6.2.8.3 Setting the upper scale value for 10 V of the analog output
You can set which leakage rates are represented by 10 V on the analog output. This
setting only comes into effect when "leakage rate linear" is selected for the
configuration of the analog output.
1
> Setup > Interfaces > I/O module > Analog scale
2
To set the upper scale value, enter a numerical value between, for example, 10
g/a and 1000 g/a using the buttons.
3
Confirm with
.
6.2.8.4 Setting up the I/O module protocol
If necessary, set the format of the I/O module protocol.
1
> Setup > Interfaces > I/O module > Protocol
ð
See also "Interface description HLD6000 (doc. no. kirb43en1)".
2
Select between "ASCII", "LD", "Normal", and "Simple".
3
Confirm with
.
6.2.8.5 Configuring digital outputs
1
> Setup > Interfaces > I/O module > PLC outputs
2
Select one of the 8 PLC outputs and allocate a function to it:
– Setpoint 1 or 2
– CAL internal active
– CAL external active
– PROOF active
– Warning
– Error
– CAL or PROOF active
–
CAL request
–
Run up
–
Sniffer button
–
Light barrier
–
Measuring
–
Standby
–
Sniffer is closed
–
Error or warning
–
CAL internal possible
3
Choose between "Normal" or "Inverse".
4
Assign additional PLC outputs to a function if necessary.
5
Confirm with
.
Содержание HLD6000
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Страница 84: ...10 Appendix INFICON 84 90 HLD6000 Operating instructions kinb43en1 05 2107 10 3 CE Declaration of Conformity ...
Страница 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Страница 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Страница 89: ......
Страница 90: ......