23 / 90
4.7 Factory settings
(For current device settings see "Parameter list [
General factory settings
Analog output upper limit
100 g/a
Display off after
1 h
Display brightness
100 %
Diagram maximum value (log.)
+3
Diagram maximum value (lin.)
20.0 g/a
Display upper limit (lin.)
300.0 g/a
Display upper limit (log.)
+3
Display unit leakage rate
g/a
Autoscale
on
Bus module address
126
Data recording
Off
Diagram of the leakage rate
Line chart
Factor user gas 1
0.0
Factor user gas 2
0.0
Factor user gas 3
0.0
Error information operator
No. and text
Error information supervisor
No., text and info
Filter change request
On
Filter change interval
40 hours
Gas of R600a sniffer line
R600a
Gas in the SMART sniffer line
R134a
Gas of R600a sniffer line
R600a
I/O module log
ASCII
Auto standby interval
2 min.
Interval for calibration request
60 min.
Calibration factor
15.0 (Calibration factor when first
commissioning a standard device).
The calibration factor cannot be reset to
factory settings. It can be changed by the
service team.)
Config. Analog output 1
Leakage rate linear
Config. Analog output 2
Leakage rate linear
Configuration PLC Output 1
Setpoint 1 (inverse)
Содержание HLD6000
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Страница 84: ...10 Appendix INFICON 84 90 HLD6000 Operating instructions kinb43en1 05 2107 10 3 CE Declaration of Conformity ...
Страница 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Страница 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Страница 89: ......
Страница 90: ......