16
Description
4.8
Technical data
4.8.1
Mechanical data
4.8.2
Physical data
4.8.3
Electrical data of the backing pump
4.8.4
Ambient conditions
4.8.5
Factory settings
The following table shows the factory settings of the control unit in the "XL Sniffer
Adapter" operating mode.
Dimensions XL Sniffer Adapter (W x H x D)
106.2 mm x 128.4 mm x 49.2 mm
Weight XL Sniffer Adapter
500 g
Gas flow
1
– High Flow
– Low Flow
1. Measured at 1 atm (1013 mbar) at sea level. The gas pressure changes with geographical height and
barometric pressure.
3000 sccm
300 sccm
Minimum detectable leak rate (MDLR)
– Helium
– High flow
– Low flow
– Forming gas (95/5)
– High flow
– Low flow
2 x 10
-6
mbar l/s
2 x 10
-7
mbar l/s
2 x 10
-6
mbar l/s
2 x 10
-7
mbar l/s
Response time
– High Flow
– Low Flow
< 1 s
< 1 s
Supply voltage
24 V DC ±10 %
Power input
4.5 A
Permissible ambient temperature
(during operation)
10 °C ... 45 °C
Permissible storage temperature
-20 °C ... 60 °C
Max. relative humidity up to 31 °C
80%
Max. relative humidity from 31 °C to 40 °C
linearly decreasing from
80% to 50%
Max. relative humidity above 40 °C
50%
Protection class
IP20
Pollution degree
II
Max. altitude above sea level
2000 m