22
Installation
4
Connect the replacement cable (200005089) with the converter of the turbo mo-
lecular pump, the MSB box (connection "2") and the fan.
5
Push the RJ45 plug labeled "XL-Adapter" into connection "A" of the XL Sniffer
Adapter.
►
Connecting the service interface:
Creating a connection between the "Service" connection on the MSB box and
connection "B" on the XL Sniffer Adapter with the supplied SUB-D cable
(200005092).
►
Connecting the sniffer interface with the MSB box:
Creating a connection between the "Sniffer" connection on the MSB box and con-
nection "C" on the XL Sniffer Adapter with the supplied M8 cable (200005091).
►
Connecting pressure sensor PSG500
Connect the data cable of the pressure sensor with connection "1" of the MSB box.
5.8
Connecting the SL3000XL sniffer line
1
Connect the end of the SL3000XL sniffer line with the front of the XL Sniffer
Adapter.
2
Fix the SL3000XL sniffer line in place so that no tensile forces are imposed on the
connection.
5.9
Start-up
5.9.1
Switching on the LDS3000
Check before the start-up:
The components of the XL Sniffer Adapter set have been installed.
The hose connections are properly connected.
The data connections and electrical connections are in place.
The backing pump is ready for operation.
►
Switching on the LDS3000.
After the start-up, the green LED on the front cover of the XL Sniffer Adapter lights
up.
5.9.2
Checking and updating the software version of the MSB box
The functions of the XL Sniffer Adapter Set are taken into consideration in the soft-
ware version 2.11 or higher.
The current software is available from the INFICON support.
The software is in the main directory of a USB stick.
1
Connect the USB stick with the USB port of control unit CU1000.