10
Description
4
Description
4.1
Function
The XL Sniffer Adapter set amends the LDS3000. With the XL Sniffer Adapter set,
even small leaks can be detected from a great distance. The best detection limit is
achieved in low flow. Switching to high flow achieves a better distance sensitivity
while the detection limit is made worse.
The flow-rate in low flow is approx. 300 sccm, in high flow approx. 3000 sccm.
Helium or forming gas (hydrogen) are used as sample gases.
4.2
XL Sniffer Adapter Set
The following graphic shows the flexible hose and the electrical connections that are
required for the connection of the XL Sniffer Adapter.
Hose connection
Electrical connection
Flange
XL Sniffer Adapter
The XL Sniffer Adapter is the interface between the electrical and the vacuum unit,
i.e. between LDS3000 and SL3000XL. It is ready for front panel installation.
SL3000XL
Fig. 1
Connection diagram of XL Sniffer Adapter set, LDS3000, backing pump, PSG500
HIGH FLOW
LOW FLOW
B
A
C
2
Service
Sniffer
LDS3000
XL Sniffer Adapter
GROSS FOREPUMP
FINE SNIFFER
SL3000XL
FINE
GROSS
XL HIGH FLOW
Forepump
FINE
1
PSG500
TMP
MSB
Filter