background image

Description

11

The sniffer line belonging to the XL Sniffer Adapter is the SL3000XL.

FINE throttle flange

The FINE throttle flange is on the "FINE Sniffer" connection and is the connection to 
the LOW FLOW connection of the XL Sniffer Adapter. It also includes the connection 
to the FINE input of the backing pump.

GROSS throttle flange

The GROSS throttle flange is on the "GROSS FOREPUMP" connection and con-
nects the backing pump with the LDS3000.

Pressure sensor PSG500

The pressure sensor PSG500 is on the second FINE connection.

TMP

"TMP" is the abbreviation for turbo molecular pump.

MSB

"MSB" is the term for the connection block.

4.3

XL Sniffer Adapter

a

Connection SL3000XL, connection for the sniffer line on the front

b

The Status LED indicates the status of the XL Sniffer Adapter. A permanently lit status LED 
shows that the XL Sniffer Adapter is supplied with voltage.

Fig. 2

XL Sniffer Adapter - Front view

XL Sniffer Adapter

a

b

Содержание 560-319

Страница 1: ...XL Sniffer Adapter Set Translation of the original operating instructions Catalog no from software version Document no 560 319 Control unit 2 11 MSB 2 11 jinxa54en1 03 ...

Страница 2: ...pplies to the software version stated on the title page Documents for other software versions are available from our sales department Reprinting translation and duplication require written confirmation from INFICON GmbH ...

Страница 3: ...flange 13 4 6 Backing pump 14 4 7 Sniffer line SL3000XL 15 4 7 1 Operating elements on the handle 15 4 8 Technical data 16 4 8 1 Mechanical data 16 4 8 2 Physical data 16 4 8 3 Electrical data of the backing pump 16 4 8 4 Ambient conditions 16 4 8 5 Factory settings 16 5 Installation 18 5 1 XL Sniffer Adapter assembling 19 5 2 Mounting the backing pump 19 5 3 Installing the FINE throttle flange on...

Страница 4: ... 6 1 Menu 24 6 2 Measuring 24 7 Maintenance 25 7 1 Indication of the next maintenance interval and total operating time 25 7 2 Maintenance message 25 7 3 Backing pump 25 7 4 Replacing the filter cartridge on the sniffer probe 25 7 5 Replacing the filter on the FINE throttle flange 26 7 6 Calibration 26 8 Decommissioning the device 27 8 1 XL Sniffer Adapter set disposal 27 8 2 Sending the XL Sniffe...

Страница 5: ... Other applicable documents 1 3 Displaying information 1 3 1 Warnings 1 3 2 Text markings Mass spectrometer module operating instructions jiqa54 Operating instructions for CU1000 jina54 Operating instructions I O module jiqc10 Protocol Descriptions jira54 Imminent threat resulting in death or serious injuries Hazardous situation resulting in potential death or serious injuries Hazardous situation ...

Страница 6: ...6 About these instructions Instruction 1 2 3 Several instructions in a fixed order Result of an action Marking Meaning ...

Страница 7: ...ith these instructions Fulfill and ensure compliance with the following regulations Intended use Generally applicable safety and accident prevention regulations International national and local standards and guidelines Additional device related provisions and regulations Use only original parts or parts approved by the manufacturer Keep this manual available at the operating site Personnel qualifi...

Страница 8: ...lation and mainte nance work The sniffer probe is made of metal and is electrically conducting Live parts should not come into contact with the sniffer probe Before starting the leak test switch off electrically operated test objects and en sure the absence of voltage for the duration of the test The unit contains electrical components that can be damaged from high electric volt age Before connect...

Страница 9: ...he XL Sniffer Adapter and the overall configuration Menu tree of the control unit CU1000 these operating instructions in a PDF file die current software for the LDS3000 Please check the scope of delivery of the product for completeness after receipt Transport Storage Always store the device in compliance with the technical data see chapter 4 8 page 16 Item Quantity XL Sniffer Adapter 1 GROSS throt...

Страница 10: ... gases 4 2 XL Sniffer Adapter Set The following graphic shows the flexible hose and the electrical connections that are required for the connection of the XL Sniffer Adapter Hose connection Electrical connection Flange XL Sniffer Adapter The XL Sniffer Adapter is the interface between the electrical and the vacuum unit i e between LDS3000 and SL3000XL It is ready for front panel installation SL300...

Страница 11: ...on the GROSS FOREPUMP connection and con nects the backing pump with the LDS3000 Pressure sensor PSG500 The pressure sensor PSG500 is on the second FINE connection TMP TMP is the abbreviation for turbo molecular pump MSB MSB is the term for the connection block 4 3 XL Sniffer Adapter a Connection SL3000XL connection for the sniffer line on the front b The Status LED indicates the status of the XL ...

Страница 12: ... the re placement cable kit d Connection B RS232 This connection transmits the signals of the sniffer line to the LDS3000 The cable is connected with the Service connection of the MSB box The cable is part of the replacement cable kit e Connection C M8 The cable is connected with the Sniffer connection of the MSB box It transmits the switching signals between low flow and high flow of the sniffer ...

Страница 13: ...dapter c Flange connection to the FINE connection of the connection block on the LDS3000 4 5 GROSS throttle flange a Flange connection to the GROSS connection of the connection block on the LDS3000 b Hose connection to the GROSS connection of the backing pump Fig 4 FINE throttle flange for the connection to the LDS3000 a b c Fig 5 GROSS throttle flange for the connection to the LDS3000 a b ...

Страница 14: ...on high flow Hose connection to the HIGH FLOW connection of the XL Sniffer Adapter b Hose connection GROSS Hose connection to GROSS throttle flange on the LDS3000 c Hose connection Fine Hose connection to FINE throttle flange on the LDS3000 d Pump outlet The pump outlet has sound absorber Fig 6 Backing pump a b c d ...

Страница 15: ...n warnings or error messages are generated The message can be acknowledged with the right button The right button can otherwise be used to switch between low flow and high flow The left button can be used for a ZERO adjustment The background display is set to ZERO by pressing the key The handle of the probe has LEDs that illuminate the area around the line The lighting behavior of the LEDs can als...

Страница 16: ...essure changes with geographical height and barometric pressure 3000 sccm 300 sccm Minimum detectable leak rate MDLR Helium High flow Low flow Forming gas 95 5 High flow Low flow 2 x 10 6 mbar l s 2 x 10 7 mbar l s 2 x 10 6 mbar l s 2 x 10 7 mbar l s Response time High Flow Low Flow 1 s 1 s Supply voltage 24 V DC 10 Power input 4 5 A Permissible ambient temperature during operation 10 C 45 C Permi...

Страница 17: ...info Viewer Number only Used trigger 1 Volume 8 Linear or logarithmic Log Time axis scale 30 s Record interval 500 ms Memory location USB Show warnings On Show value On Parameters Factory setting Filter I Filter Flow control low LOW FLOW Flow button to control on Gas concentration H2 5 Capillary surveillance Pressure capillary blocked low flow Pressure capillary broken low flow Pressure capillary ...

Страница 18: ...ng fixture must be provided by the operator Sufficient ventilation must be ensured for a closed housing Fig 8 Assembly of the components h g f e d c b a a Pressure sensor PSG500 e Connections on the MSB box b Turbo molecular pump f Control panel CU1000 c FINE throttle flange g XL Sniffer Adapter d GROSS throttle flange h Backing pump ...

Страница 19: ...NF backing pump 5 3 Installing the FINE throttle flange on the connection block 1 Remove the cover from one of the FINE ports of the LDS3000 connection block 2 Remove the blank flange from the FINE port 3 Remove the protective cover on the FINE throttle flange 4 Attach the FINE throttle flange with the centering ring on the FINE port and attach with the clamping ring Fig 9 Dimensions of the adapte...

Страница 20: ... backing pump PSG500 auf Seite 10 Connect the HIGH FLOW output of the XL Sniffer Adapter with XL HIGH FLOW input of the backing pump Connect the LOW FLOW output of the XL Sniffer Adapter with the filter of the FINE throttle flange Screw the hose installed on the XL Sniffer Adapter onto the filter on the FINE throttle flange with a luer lock connection Connect the FINE throttle flange on the LDS300...

Страница 21: ...ase and pull off the SUB D plugs on the converter of the turbo molecular pump and on the MSB box connection 2 3 Release and pull off the plug on the fan Fig 10 Cable tie to open Fig 11 The three plugs of the cable to replace ...

Страница 22: ...0XL sniffer line 1 Connect the end of the SL3000XL sniffer line with the front of the XL Sniffer Adapter 2 Fix the SL3000XL sniffer line in place so that no tensile forces are imposed on the connection 5 9 Start up 5 9 1 Switching on the LDS3000 Check before the start up The components of the XL Sniffer Adapter set have been installed The hose connections are properly connected The data connection...

Страница 23: ...ate Control unit The display shows information on the current and the new software version 3 Check the version information 4 Press Start to begin the update and confirm the prompt with OK 5 9 4 Maintenance interval of the backing pump activation Activate the maintenance interval for the commissioning Access Integrator 1 Select the main menu Authorizations Integrator Maintenance Type of maintenance...

Страница 24: ...hrough breathing air forklift and the like that could influence the background of the measurement If helium is measured the helium concentration in the measuring environment must be kept as low as possible The display on the handle shows if and how much gas is measured see Operating elements on the handle auf Seite 15 Leaking gas can be indicated with flashing LEDs an acoustic signal and vibration...

Страница 25: ... Once a maintenance interval has been reached a maintenance message is dis played 7 3 Backing pump The backing pump must be serviced regularly refer to the documentation of the backing pump 7 4 Replacing the filter cartridge on the sniffer probe The replacement interval depends on the ambient conditions The filter cartridge must generally be replaced every 500 to 1500 operating hours The filter ca...

Страница 26: ...s no warning tighten the cap nut and if that is not successful check the seat of the filter 3 Tighten the plastic cap on the end the sniffer probe again 4 Calibrating the LDS3000 7 5 Replacing the filter on the FINE throttle flange The replacement interval depends on the ambient conditions The filter must gener ally be replaced every 500 to 1500 operating hours 7 6 Calibration Calibration is neces...

Страница 27: ...local and regional environmental and safety reg ulations 8 2 Sending the XL Sniffer Adapter Set to INFICON The declaration of contamination is a legal requirement and serves to protect our em ployees INFICON sends devices which are sent without a completed declaration of contamination back to the sender Danger to health Contaminated devices could endanger the health of INFICON employees Fill in th...

Страница 28: ...28 Decommissioning the device 8 3 Declaration of Contamination ...

Страница 29: ...Declaration of Incorporation 29 9 Declaration of Incorporation ...

Страница 30: ...498 D 50968 Cologne Germany UNITED STATES TAIWAN JAPAN KOREA SINGAPORE GERMANY FRANCE UNITED KINGDOM HONG KONG Visit our website for contact information and other sales offices worldwide www inficon com Dokument jina54en1 03 1512 ...

Отзывы: