Description
15
4.7
Sniffer line SL3000XL
4.7.1
Operating elements on the handle
The display of the handle displays part of the information of the main display.
The leak rate is shown as a bar graph and displayed numerically. The unit of mea-
surement is the same as in the main display.
The display also shows the gas type and the tracer gas concentration. If the XL Sniff-
er Adapter is operated in the high flow operating mode, then the gas type display has
a dark background.
The display shows the messages when warnings or error messages are generated.
The message can be acknowledged with the right button.
The right button can otherwise be used to switch between low flow and high flow.
The left button can be used for a ZERO adjustment: The background display is set
to ZERO by pressing the key.
The handle of the probe has LEDs that illuminate the area around the line.
The lighting behavior of the LEDs can also signal that a trigger has been exceeded,
see "Settings for the XL Sniffer Adapter" n the operating instructions for the
LDS3000.
Fig. 7
Sniffer line SL3000XL
Risk of eye damage
LEDs generate a bundled light that can damage your eyes.
►
Do not look into the LEDs from a short distance or for longer periods of time.