Installation
23
2
Select the main menu:
"Functions > Data > Update > MSB"
The display shows information on the current and the new software version as
well as on the boot loader.
3
Check the version information.
4
Press "Start" to begin the update and confirm the prompt with "OK".
5
If the system issues warning 106, confirm this with "C".
5.9.3
Updating the software of control unit CU1000
The software is included in two files named
Handset_IFC_Vx.xx.xx.exe and Handset_IFC_Vx.xx.xx.key.
The new files are in the main directory on a USB stick.
1
Connect the USB stick with the USB port of the control unit.
2
Select the main menu:
"Functions > Data > Update > Control unit"
The display shows information on the current and the new software version.
3
Check the version information.
4
Press "Start" to begin the update and confirm the prompt with "OK".
5.9.4
Maintenance interval of the backing pump activation
Activate the maintenance interval for the commissioning.
Access = Integrator
1
Select the main menu:
"Authorizations > Integrator > Maintenance > Type of maintenance"
2
Select "Membrane"
3
"Maintenance work > Completed" confirm with "OK"
The maintenance interval is activated.
5.9.5
Maintenance message activation
To optionally receive a maintenance message when the maintenance interval has
been reached, the must be activated.
Access = Integrator
1
Select the main menu:
"Authorizations > Integrator > Maintenance> Maintenance message"
2
Select "TMP + Membrane" and confirm with "OK"
The maintenance message is activated.