O P E R A T I O N S
Functions and Features
38
LDC-3700C Series
C H A P T E R
2
instrument is restarted and the front panel parameters are reconfigured to the new
parameters saved under that bin number.
Using the Laser Compliance Voltage Adjustment
For applications where you need to have critical protection of the laser we
recommend the following. Using a 1
resistor or other non-critical load, set the
LASER voltage limit as described in Chapter 2.
Then with the LASER output on, slowly increase the LASER current and watch
the LASER voltage measurement. (Press the
LASER DISPLAY V
push button to
see the voltage). At about 0.25 Volts below the LASER voltage limit value, the
VOLTAGE LIMIT warning indicator will become lit on the
LASER display
.
Continue to adjust the current up until the voltage limit is reached and the output is
shut off. Record the voltage at which the LASER output shuts off. It should be
within the specification of the LASER voltage limit accuracy. However, there is
typically some offset error. For example, if the LASER voltage limit is set to 5.0
Volts, the actual voltage where it shuts off may be 4.9 Volts. This error term of 0.1
Volts should be considered when setting the LASER compliance voltage limit.
The closer the LASER voltage limit shut off point is to the operating voltage of
your laser, the faster the circuit will work in the event of an open circuit. Some
experimentation may be necessary for optimum results.
Using the LDC-3700C Controller’s Trigger Function
For applications where you need to synchronously initiate a measurement task
from a remote instrument with the LDC-3700C Controller, the controller offers a
trigger output signal. The TTL pulse is initiated with any change in set point of the
laser current source and after the analog output has settled. A typical application
for utilization of this feature is laser characterization tasks where an L-I curve is
generated. For a programmed step in laser current, a light measurement can be
triggered for each step in laser current. The trigger output is enabled only in
remote mode via the IEEE 488.2 GPIB Interface.
Note:
The minimum step change time required for a corresponding trigger pulse is 20 ms.
Care should be used with respect to the timing of any set point commands in relation to the
actual hardware function. The INC and DEC commands have a fixed minimum step time of
20 ms which allows for the hardware initiation of a set point change and the settling of the
output. The LDI command does not. With this command, some trigger pulses may be
missed if the program step time is less than the 20 ms minimum.
A one shot trigger pulse will occur on power up of the instrument due to the states
of the processor I/O.
L AS E R D IS P L AY
I
I
P D
P
P D
V
S ET
Содержание LDC-3700C Series
Страница 2: ......
Страница 8: ...L I S T O F FIG U R E S vi LDC 3700C Series...
Страница 10: ...L I S T O F TA B L E S viii LDC 3700C Series...
Страница 78: ...RE M O T E O PE R A T I ON Command Syntax 62 LDC 3700C Series C H A P T E R 3...
Страница 150: ...C O M M A N D RE F E R E N C E Command Reference 134 LDC 3700C Series C H A P T E R 4...
Страница 176: ...C A L IB R A TI O N A N D TR OU BL E S HO O T IN G Troubleshooting 160 LDC 3700C Series C H A P T E R 5...