C O M M A N D R E F E R E N C E
Command Reference
132
LDC-3700C Series
C H A P T E R
4
The TERM command allows the programmer to change the default (IEEE488.2
standard) terminator to include the carriage return.
Parameters
An <nrf value>, 0 = FALSE, non-zero = TRUE
Notes
An altered terminator will be in the form <CR><NL><^END>. This technically takes the
LDC-3700C Series Controller out of IEEE488.2 specification, but may be done for
convenience when using non standard GPIB controllers. This termination will be sent
with all output until the "TERM 0" command is sent or the 3700C Series Controller is
powered off.
Examples
"Term 1" -temporarily sets <CR><NL><^END> as the output terminator.
"Term?" -response: 0, means that the <NL><^END> terminator (IEEE488.2 standard) is
being used.
The TERM? query allows the programmer to determine whether the default (IEEE488.2
standard) terminator has been altered to include a carriage return.
Notes
An altered terminator will be in the form <CR><NL><^END>. This termination will be
sent with all output until the "TERM 0" command is sent or the LDC-3700C Series
Controller is powered off.
Examples
"Term?" -response: 1, means that the <CR><NL><^END> terminator is temporarily
being used.
"Term?" -response: 0, means that the <NL><^END> terminator (IEEE488.2 standard) is
being used.
The TIME? query allows the programmer to determine how much time has passed since
the LDC-3700C Series Laser Diode Controller was last powered up.
Notes
The TIME clock is independent of the TIMER clock.
Examples
"Time?" -response: 1:02.36, means that 1 minute and 2.36 seconds have passed since
the LDC-3700C Series Laser Diode Controller was powered up.
"TIME?" -response: 32:00:76, means that 32 minutes and 0.76 seconds have passed
since the LDC-3700C Series Laser Diode Controller was powered up.
TERM
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
TERM?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
TIME?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
Содержание LDC-3700C Series
Страница 2: ......
Страница 8: ...L I S T O F FIG U R E S vi LDC 3700C Series...
Страница 10: ...L I S T O F TA B L E S viii LDC 3700C Series...
Страница 78: ...RE M O T E O PE R A T I ON Command Syntax 62 LDC 3700C Series C H A P T E R 3...
Страница 150: ...C O M M A N D RE F E R E N C E Command Reference 134 LDC 3700C Series C H A P T E R 4...
Страница 176: ...C A L IB R A TI O N A N D TR OU BL E S HO O T IN G Troubleshooting 160 LDC 3700C Series C H A P T E R 5...