C O M M A N D R E F E R E N C E
Command Reference
88
LDC-3700C Series
C H A P T E R
4
The LASer:ENABle:OUTOFF? query returns the value of the status outoff enable register
of the LASER operations.
-where the response is the sum of the following:
Notes
The enabled LASER events can be set by using the LASer:ENABle:OUTOFF command.
The LASER output status can be monitored by the LASer:EVEnt? query.
Examples
"LAS:ENAB:OUTOFF?" -response: 2049, means that TEC High Temperature Limit and
Current Limit Conditions will cause the LASER output to be turned off.
"Radix Hex; Las:Enab:Eve?" -response: #HE0B, means that all of the above conditions
will cause the LASER output to be turned off.
The LASer:EVEnt? query returns the value of the status event register of the LASER
operations.
-where the response is the sum of the following:
Notes
The LASER conditions that are reported in the status byte can be set by using the
LASer:ENABle:EVEnt command.
The LASER event status is only cleared when the event status is read or by the *CLS
command, while the condition status is constantly changing.
Examples
"LAS:EVE?" -response: 513, means that the LASER output tolerance changed and
current limit events have occurred since the last LASer:EVEnt? query.
"Radix Hex; Laser:Event?" -response: #H88, means that the LASER Power limit and
Open circuit events have occurred since the last LASer:EVEnt? query.
LASer:ENABle:OUTOFF?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
1 - LASER Current Limit
256 - N/A
2 - LASER Voltage Limit
512 - Output is Out of Tolerance
4 - N/A
1024 - TEC Output Off
8 - Power Limit (With Output On)
2048 - TEC High Temperature Limit Condition
16 - N/A
4096 - N/A
32 - N/A
8192 - N/A
64 - N/A
16384 - N/A
128 - N/A
32768 - N/A
LASer:EVEnt?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
1 - LASER Current Limit
256 - Output is Shorted
2 - LASER Voltage Limit
512 - Output Changed to be In/Out of Tolerance
4 - N/A
1024 - Output On/Off State Changed
8 - Power Limit
2048 - Measurement Ready
16 - Interlock Disabled
4096 - N/A
32 - N/A
8192 - N/A
64 - N/A
16384 - N/A
128 - Open Circuit
32768 - N/A
Содержание LDC-3700C Series
Страница 2: ......
Страница 8: ...L I S T O F FIG U R E S vi LDC 3700C Series...
Страница 10: ...L I S T O F TA B L E S viii LDC 3700C Series...
Страница 78: ...RE M O T E O PE R A T I ON Command Syntax 62 LDC 3700C Series C H A P T E R 3...
Страница 150: ...C O M M A N D RE F E R E N C E Command Reference 134 LDC 3700C Series C H A P T E R 4...
Страница 176: ...C A L IB R A TI O N A N D TR OU BL E S HO O T IN G Troubleshooting 160 LDC 3700C Series C H A P T E R 5...