
C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Troubleshooting
04_09
LDC-3700C Series
159
C H A P T E R
5
E-517
Calibration for Laser Diode current must have the output on to start.
E-518
Calibration for the Monitor Diode must have the output on and the responsivity set
to zero to start.
E-519
Setting a measurement is only valid during the calibration phase for that
measurement. User has tried to calibrate a measurement without first entering the
required calibration mode.
E-520
User cannot change the Laser Current set point while operating in a calibration
mode for another measurement.
Table 5.5
Error Messages
Error Code
Explanation
Содержание LDC-3700C Series
Страница 2: ......
Страница 8: ...L I S T O F FIG U R E S vi LDC 3700C Series...
Страница 10: ...L I S T O F TA B L E S viii LDC 3700C Series...
Страница 78: ...RE M O T E O PE R A T I ON Command Syntax 62 LDC 3700C Series C H A P T E R 3...
Страница 150: ...C O M M A N D RE F E R E N C E Command Reference 134 LDC 3700C Series C H A P T E R 4...
Страница 176: ...C A L IB R A TI O N A N D TR OU BL E S HO O T IN G Troubleshooting 160 LDC 3700C Series C H A P T E R 5...