O P E R A T I O N S
Operating the TE Controller from the Front Panel
36
LDC-3700C Series
C H A P T E R
2
TEC Error Indicators
The LDC-3700C Series Laser Diode TEC Controllers indicate general TEC
operational error conditions. Each error condition results in a controller action as
shown in the following table.
Table 2.3
TEC Error Indicators
Error Condition
Action
Temperature limit
TEMP LIMIT light flashes at 1 Hz
Open sensor
Output off, SENSOR OPEN indicator LED flashes at 1 Hz
TE Current limit
TE CURRENT LIMIT light flashes at 1 Hz
TE Module Open
TE MODULE OPEN indicator LED flashes at 1 Hz (not in
effect when booster source is used)
TE Compliance Voltage Limit
TE CURRENT LIMIT light flashes at 2 Hz