LDC-3700C Series
13
C H A P T E R
2
O
PERATIONS
This chapter introduces you to the operation of the LDC-3700C Series Controllers.
It offers instructions for connecting your laser to the current source and
temperature controller and describes powering up the instrument. This chapter
also contains step by step procedures that teach you how to operate your
controller in Constant Current Mode, Constant Power Mode, and to operate the
temperature controller in Constant Temperature Mode. ILX recommends that you
review the contents of this chapter at a minimum before operating your new
controller.
Applying Power to your LD Controller
To turn on the LDC-3700C Controller, rotate the key in the "POWER" section of
the front panel from "O" to "I". This action initiates the power on sequence. If the
LDC-3700C does not appear to turn on, verify that it is connected to line power. If
line power is not the problem, remove the power cord and check the line power
fuse.
Note:
The instrument contains a laser current source that has remote interlock capability.
This allows you to connect the interlock circuit to an external switch for safety. If the
interlock is removed, the laser current output is disabled. Refer to the laser module manual
for details about the interlock function.
The instrument is also equipped with a laser enable key switch. This ensures that the laser
current can be turned on only if the key is inserted and in the On position.
The Power On Sequence
During the power-up sequence, the following takes place. For about three
seconds all indicators light up, and all of the 7-segment displays indicate "8". Then
all lamps are turned off for three seconds. Next, the firmware version displays in
the LASER display section followed by the TEC sensor switch position displayed
on the TEC display. Then, a self-test is performed to ensure that the instrument's
hardware and processor are communicating. After this test, the instrument is
ready to operate, and is configured to the state it was in when the power was last
shut off.
Содержание LDC-3700C Series
Страница 2: ......
Страница 8: ...L I S T O F FIG U R E S vi LDC 3700C Series...
Страница 10: ...L I S T O F TA B L E S viii LDC 3700C Series...
Страница 78: ...RE M O T E O PE R A T I ON Command Syntax 62 LDC 3700C Series C H A P T E R 3...
Страница 150: ...C O M M A N D RE F E R E N C E Command Reference 134 LDC 3700C Series C H A P T E R 4...
Страница 176: ...C A L IB R A TI O N A N D TR OU BL E S HO O T IN G Troubleshooting 160 LDC 3700C Series C H A P T E R 5...