Connected power supply
The generator beeps and the
100% FULL battery symbol is displayed.
The battery is charged:
disconnect the power
supply
.
The device switches OFF when the power
supply is disconnected.
5.5
Battery efficiency
The efficiency of the battery is affected by the correct use and wear of the battery.
If the battery does not allow 4 consecutive hours of treatment, daily treatment can be performed using the external
power supply.
When the display indicates that the battery is running low, the battery symbol turns red and a plug symbol appears
next to it. Connect the power supply to the generator and to the mains socket,
keeping the generator switched on
:
the device beeps and vibrates, indicating that the charger has been connected to the generator. The plug symbol
disappears and a flash is displayed in its place, indicating that charging is in progress. The device continues to
administer the treatment while charging the battery. At the end of the treatment, leave the generator switched off
and connected to the power supply until the battery charge is complete.
5.6
Treatment times
The user must undergo the treatment for the number of days indicated by the prescribing doctor.
Clinical studies have shown that
the I-ONE treatment is effective if performed for 4 hours a day;
in any case, there
are no problems or side effects due to overdosing.
It is good practice to perform the daily treatment in a single session; however, it is possible to split the treatment
time into several daily applications of
no less than 2 hours
. The absence of side effects means that the treatment
can be carried out even whilst sleeping.
5.7
Useful Tips
•
To make using the device easier, it is recommended to leave the coil connected to the generator to avoid
having to keep connecting it for each new treatment session.
•
It is recommended that the battery be charged every day after the treatment is completed, so that the next
treatment can be fully performed. Please note that it is also possible to carry out the treatment using the
mains power by connecting the generator to the external power supply.
•
The parts of the device that may come into contact with the skin do not normally cause any allergic reaction.
Although the coil's covering material is hypoallergenic and biocompatible, it is recommended not to place
the coil in direct contact with the skin, but to place it on a light garment, especially if there is any reddening
or irritation in the area of application.
•
Clean the coil regularly, taking care to disconnect the coil from the generator using neutral detergents.
•
Any wear on the coil coating due to use does not affect the effectiveness of the treatment. However, in the
event of a loss of some of the coating, the coil must be replaced.
•
When using the coil under heavy blankets, the coil surface may become overheated. If this is the case,
perform the treatment without covering the coil.
•
In any case, the generator must not be covered to allow ventilation whilst it is in operation.
•
Using the device in environments with a temperature above 30°C, although possible, could cause the coil
surface to slightly overheat and this may be uncomfortable for the user.
•
When charging or operating via the mains, it is normal for the battery to get hot; for this reason, the
generator should not be placed in direct contact with the body when charging or operating via the mains.
•
The elastic band can be washed just like any other garment.