14
Mobil-O-Graph
®
, operating instructions, Rev. 5.1 10/2015, GmbH
3. Product description
You can set the protocols manually by holding down the Day/Night and then pressing the Event key. Use the
Start/Stop key to change the protocol, and confirm with the Event key.
Protocol
Day time
Night time
Measurements
per hour
Acoustic signal
Measurements
shown on display
1
08:00
00:00
23:59
07:59
4
2
YES
NO
YES
2
08:00
23:00
22:59
07:59
4
1
YES
NO
YES
3
07:00
22:00
21:59
06:59
4
2
YES
NO
NO
4
08:00
00:00
23:59
07:59
4
2
YES
NO
NO
5
18:00
10:00
09:59
17:59
4
2
YES
NO
YES
6
07:00
00:00
23:59
06:59
4
2
YES
NO
YES
7
06:00
23:00
22:59
05:59
4
2
YES
NO
NO
8
07:00
09:00
00:00
08:59
23:59
06:59
6
4
2
YES
YES
NO
YES
9
09:00
08:59
30
NO
YES
10
08:00
07:59
30
YES
NO
11
08:00
00:00
23:59
07:59
4
2
YES
NO
YES
Table 3.0 Protocol list
Note:
Protocols 1, 2 and 10 are provided as defaults, but they can be modified using the analysis soft-
ware (see 4.8, prepare measurement device). Protocol 5 is suitable for nighttime activity (night shift).
Protocol 9 is the Schellong-test.
Protocol 10 transmits the values automatically after each measurement to your doctor’s PC via
Bluetooth, or to a database via mobile communications such as a mobile phone or modem (precon-
figured by I.E.M.; see 4.8.1, Setting the measurement protocol). Only preconfigured mobile phones
or modems can send the data via GPRS as a block of data.
Protocol 11 is only available for the Mobil-O-Graph
®
with integrated 24-h-PWA and the HMS CS as
of version 4.0. Here, the intervals for blood pressure monitoring and 24-h-PWA can be configured
independently
3.4.6
Putting on the monitor and starting measurement
Figure 3.2 Putting on the cuff
Put the equipment pouch on the patient. By varying the length of
the strap you can use it either as a hip belt or a shoulder strap.
Alternatively, use a normal belt that matches the patient’s clothing.
Now fit the cuff on the patient; correct cuff fit is extremely important
for accurate measurement (see
Figure 3.2
).
The cuff can also be worn over the shirt or blouse. We recommend
that the cuff be fitted on the bare upper arm. In this case, bring the
tube out between the buttons of the shirt or blouse, around the back
of the neck to the ABP monitor on the right-hand side.
3.4.6