59
Mobil-O-Graph
®
, operating instructions, Rev. 5.1 10/2015, GmbH
4.
H
ypertension
M
anagement
S
oftware
Client-Server
4.11.14 Patient report (Day Average)
In the case of PWA measurements as part of 24h ABPM values, the patient report displays the average of select-
ed haemodynamic information determined during the measurement period.
In the case of PWA measurements in the doctor’s office, the patient report displays haemodynamic information
determined from the single measurement.
To access the patient report (day average), click on the
Patient report (Day Average)
tab
.
The
Patient report (Day Average)
tab is divided into four areas:
Peripheral blood pressure
:
The peripheral blood pressure area displays the peripheral systolic and peripheral diastolic blood
pressure measured.
Central blood pressure
:
The central blood pressure area displays the central systolic blood pressure calculated.
Arterial stiffness
:
The arterial stiffness area displays the PWV (pulse wave velocity) measured for large vessels and the
augmentation index at a heart rate of 75 1/min (AIx@75) in the case of small vessels.
Vascular age
:
The vascular age area shows the vascular age of the patient determined from the haemodynamic
information. The vascular age display can be set in the Hypertension Management Software settings.
Central blood pressure and calibration:
Currently, calibration is used when calculating the central aortic systolic blood pressure. As a rule, peripheral
systolic blood pressure is used. Calibration can also be used according to the MAP measured (mean arterial
pressure, corresponds to the maximum oscillometric amplitude).
Scientific findings published recently prove for the first time that calibration according to the measured MAP is
more predictive than other comparable methods.
To use the
MAP-C2 calibration
for calculating the aortic central systolic blood pressure, in the menu bar, click
on
Settings -> Extras
and enable
MAP-C2-calibration for PWA.
4.11.14