62
Mobil-O-Graph
®
, operating instructions, Rev. 5.1 10/2015, GmbH
4.
H
ypertension
M
anagement
S
oftware
Client-Server
►
Click on
export patient
►
The
export
window appears
►
Click on yes
Already opened patients can also be exported using the
symbol and synchronized using the
symbol.
After the import or export of patients, the patients authorized for e-health are marked with a small
in the
patient list and thereby authorized for synchronization with the e-health server.
If you open such a patient, all new telemetric measurements will be automatically polled from the e-health
server.
If you want to reverse the clearance of a patient on the e-health server, click
.
4.13 Practice Monitoring
The practice monitoring should relieve the practice, increase therapy quality and improve patient comfort. The
Mobil-O-Graph can be carried by the patient in the practice, e.g. in the waiting room, and the test series is trans-
ferred directly via Bluetooth to a practice computer. Each measurement is transferred wirelessly to the HMS CS
and can be analyzed by the Doctor directly
You can use the practice monitoring for:
■
intervention during high pressure crises.
■
minimum invasive intervention (Warning: The system does not provide an alarm function!)
4.13.1 Prepare Mobil-O-Graph
®
for the practice monitoring
The
Bluetooth interface
of the Mobil-O-Graph is used for the practice monitoring. If you have never worked
with the Bluetooth interface, then please follow the instructions under 4.7.2 Connection via Bluetooth® for the
configuration of the interface.
►
Select
protocol 10 for the practice monitoring
(see 4.8.1 specify measurement protocol).
►
Choose the time interval 30, 20, 15, 12... measurements per hour.
►
Attach the measuring device to the patient (fasten the cuff and connect to the device).
►
To ensure that the device functions as desired, push
for the start of a manual measurement. A successful
measurement is a condition for the activation of the protocol!
►
Wait for the first automatic measurement.
4.13