72
Mobil-O-Graph
®
, operating instructions, Rev. 5.1 10/2015, GmbH
4.
H
ypertension
M
anagement
S
oftware
Client-Server
4
►
Click
.
The window disappears.
4.16.8 Extras
Here you can set standard procedures for:
■
Calculation of mean (of all single measurements or the hourly mean values (HM)
■
Calculation of mean arterial pressure (measured MAP or calculated MAP)
■
which measurement analysis will be displayed after transmission of measurements from the device to HMS CS
CS (Table of measurements or graph).
■
Whether the Patient list will be shown on starting the program, or an „empty” control window appears.
■
Whether you are using Bluetooth.
►
In the menu bar, click
settings
and then
Extras
.
The window
Extras
appears.
►
Establish your required settings.
►
Click
.
The window disappears.
4.16.9 GDT settings
GDT is a data exchange format used by General Practitioners in the German healthcare system. The GDT
interface is used to transmit data independently of the type of system between medical measurement devices
and practice IT systems.
GDT settings are required for the automatic transfer of patient data between your practice’s computers and
the HMS CS. If the settings are correct, the HMS CS can be started from your practice’s software, and pati-
ent data imported directly.
►
In the menu bar, click
Settings
and then
GDT settings
.
The window GDT settings appears.
►
Click Selection. Here you can set the directory shared between the HMS CS and your practice’s computer sys-
tem. The HMS CS and practice IT system must have the same setting for the directory. By preference, the pro-
gram directory of the HMS CS should be set.
4.16.8