Fig 51 shows the result screen of a successful Web test.
Frame
Size
and
Frame Fill
are fixed, and stated for reference only. The
Information rate,
IR (Mb/s)
,
is variable and dependent on the
number of sessions you have entered at set up.
Tx
states the
number of frames transmitted, the green indicator bar confirms
that 100% of the frames were sent.
Rx
states the number of
frames received, the green indicator bar confirms that 100% of
the transmitted frames were received. As all transmitted frames
were received,
Error
states 0 and its associated indicator bar
remains uncolored and shows 0%.
Fig 51
Video
Fig 52 shows the result screen of a successful Video test.
Frame
Size
and
Frame Fill
are fixed, and stated for reference only. The
Information rate,
IR (Mb/s)
, is variable and dependent on the
number of streams entered and the definition chosen at set up.
Tx
states the number of frames transmitted, the green indicator
bar confirms that 100% of the frames were sent.
Rx
states the
number of frames received, the green indicator bar confirms that
100% of the transmitted frames were received. As all transmitted
frames were received,
Error
states 0 and its associated indicator
bar remains uncolored and shows 0%.
Fig 52
CCTV
Fig 53 shows a CCTV performance test in progress.
Frame Size
and
Frame Fill
are fixed, and stated for reference only. The
Information rate,
IR (Mb/s)
,
is variable and dependent on the
Resolution, CODEC and number of cameras chosen at set up.
Tx
shows that, so far, 72,371 frames have been transmitted
which is 92% of the total frames required to be sent in this test.
Rx
shows that only 62,405 of the transmitted frames have been
received (79% of the total sent so far).
Error
shows the number
of errored (unreturned) frames (Approx. 12%).
When the number of errored frames is
≥ the Threshold (Frames)
that you entered at set up, the test is failed.
Fig 53