DOC E10523 Rev.D
Interactive Circuits And Systems Ltd. 2000
7
3.3 Clock/Trigger Options
The ICS-130 offers a number of clock and trigger options. The card has an internal
sampling clock provided by a programmable oscillator giving a resolution of less than
250Hz at the output rate over the entire 1.0 MHz to 19.2 MHz range. Alternatively, an
external sampling clock applied at pin 25 of the P4 front panel connector may be used. In
this case, the clock must be at the oversampling rate (either 16x or 32x the required output
rate).
The trigger can be programmed to be internal (i.e. software controlled) or external. If using
the external trigger, the user should supply a positive-going TTL pulse on pin 23 of the P4
connector; the pulse must be at least one sample period long. The trigger is internally
synchronized to the sampling clock by the ICS-130. Acquisition starts with the first valid
data word after the application of the trigger.
Both external clock and trigger signals must conform to standard TTL levels and drive
capability. The relevant connections are listed in section 6.3.
3.4 Data Path Selection
The ICS-130 provides three basic options for reading acquired data. The destination for the
ADC data can be VMEbus, FPDP, or P2. One or more of these interfaces may be enabled
at a time.
3.5 Modes of Operation
Three modes of operation are provided in the ICS-130 design:
- capture mode without pre storage,
- capture mode with pre storage,
- and continuous mode, which has no pre storage option.
In the capture mode without pre storage, data is acquired for a programmable number of
samples following the application of each trigger. The maximum number of samples that
can be stored for all active channels is 1048576. Because the size of the memory buffer,
and the count of samples acquired are both programmable, multiple capture acquisitions
may be stored in the ICS-130 memory.
Important note: When using capture mode without pre-trigger storage, the memory
buffer length programmed must always be an integral number of acquisition counts.