2-5
2
GPIB Bus Cable
AC Adapter
4867
GPIB Port
PC Computer
Figure 2-2 4867 Configuration Setup
3.
Apply power to the unit. The PWR LED will come on while the unit
performs its self test. When the test is done, the unit will blink its
current address setting before turning on the RDY LED.
4.
Double click on the GPIBkybd icon on your desktop or navigate to
the Programs folder to launch the GPIBkybd Program. The
GPIBkybd window will open and the program will automatically
find the unit at its GPIB address. (New units are factory set to GPIB
primary address 4)
5.
Verify communication with the unit by entering *IDN? in the
Device Command window and clicking the Send button. The unit's
IDN message will be displayed in the Device Response window.
6.
At this point you can enter the SCPI command and value to set or
control the desired IO function. Refer to the default settings in
Table 1-2. You can query a SCPI setting by ending the command
with a question mark. When you have all of the functions running
correctly, you can set them to the desired power turn-on values and
then save the settings with the *SAV 0 command.
Watch the red ERR LED closely when manually entering commands.
If it comes on, you made a command error. Clear the ERR LED by
querying the ESR Register (*ESR?) or with the clear command
(*CLS) before resending the unit the command.