5-8
5
5.4
FACTORY DEFAULT EMERGENCY RECOVERY
If the unit's configuration gets into an unknown setting, it can be restored
to the factory default configuration with the following procedure:
1.
Turn the unit's power off.
2.
For 4867 and 2367s, place a jumper on W1 or short out the two posts
on W1. For 4807 and 2307s, place a jumper on W4 or short out the
two posts on W4.
3.
Turn the unit's power on. Wait until the LEDs stabilize and the unit
has finished its save procedure (takes approximately 10 seconds).
4.
Remove the jumper and operate the unit normally. All parameters
should have been restored to the settings listed in Table 1-2 or Table
1-3.
5.
Adjust the unit's configuration for your application and do a *SAV
0 command to save the values. Do a CAL:DATE command with
the current date.