IceCure
™
Medical Ltd.
DSR3200000 rev. G
ProSense
™
Confidential
User Manual
94
Return to first page U.S User Manual
•
The skip icon enables the user to skip from the current step in the cycle to the
next step as preprogrammed.
•
Within a given protocol, if you want to increase/decrease the preprogrammed
time of a cycle when the cycle has already started, use the +/- icons on the left
side of the screen to add/subtract 15 seconds with each push.
•
At the end of the last freeze step in automatic mode, the EXTRACTION step will
begin and the EXTRACTION screen will be displayed.
•
You can move from the automatic mode AT ANY TIME by pressing the Manual
mode button on the screen, without interrupting the procedure. When in
Manual, you can’t go back to Automatic mode
.
o
If you choose to do so, proceed as explained in section
“
Manual freeze
mode
”
: press the ACTION button on the straight cryohandle to move from
freeze to THAW and monitor the process under ultrasound or other
imaging modality.
Figure 86: Automatic Mode screen: press on Start to begin the freeze cycle.