5.2.3
Switching on the
ProSense™
cryoablation system
Before operating the system, make sure the following conditions are in place:
•
Ultrasound or other appropriate imaging system is available for monitoring the
medical procedure. In case of open surgery or superficial cryotherapy, no external
imaging device is required.
•
The mechanical power switch is OFF.
•
The cryohandle, hose and cradle are all clean and dry.
To Switch on the
ProSense™
cryoablation system:
To switch the system ON, turn ON the mechanical button at the back of the
system, the touch screen will turn on and the following screen will appear:
Figure 42: The system is loading screen
Wait for the Main Menu screen to load.
Make sure the time is set according to local time zone before executing a cryotherapy
procedure, as explained in 5.2.5.4.