Illustrations of potential functional failures will appear on screen. Please refer to
these illustrations while inspecting the cryoprobe.
Figure 71: Functional test - visual inspection screens
•
If a functional problem occurs or there is any unusual appearance (such as frost
on the plastic cover near the cryohandle, foam or any unusual appearance) and it
is not represented by one of the Illustrations, press CANCEL. Follow system
instructions until you are required to safely remove the cryoprobe from the
cryohandle. You will then be returned to the Main Menu screen. To start again,
check that the Dewar is full and replace the cryoprobe. If the problem recurs, do
not proceed with treatment. Turn OFF the system and contact IceCure Medical.
•
If a functional problem occurs as represented by one of the on-screen images,
press on the relevant image.