IceCure
™
Medical Ltd.
DSR3200000 rev. G
ProSense
™
Confidential
User Manual
22
Return to first page U.S User Manual
2.4.2
Opening the Dewar compartment
Caution
Never open the Dewar compartment while running a cryoablation
procedure. Do not use the system while the Dewar compartment is open.
Warning
Do not transfer a Dewar containing Liquid nitrogen unless it is covered with
its designated lid.
Caution
Follow the specific instructions on how to fill and transfer a Dewar as
described in section 5.2.6.1
–
Dewar.
2.4.3
Voltage and power ratings
Caution
Verify that the
ProSense™ cryoablation system
complies with the local
voltage line.
2.4.4
Liquid nitrogen Safety
Warning
Liquid nitrogen may cause, if there is not sufficient ventilation in the room,
serious injury or burn if handled improperly. Local laws and safety rules
regarding the maintenance and handling of liquid nitrogen Dewars should
always be observed. Maintenance of liquid nitrogen Dewars should be
performed by authorized personnel only.
Warning
To avoid spillage and cold burns, do not transfer a Dewar containing liquid
nitrogen unless it is covered with its designated lid.
Warning
Do not use a liquid nitrogen Dewar if it is damaged.