Do not force removal of the cryoprobe from the tissue as it might increase
the risk of tissue damage. Continue the EXTRACTION step or wait for passive
THAW until the cryoprobe can be withdrawn easily.
Warning
Before removing the temperature sensor from the tissue, make sure the
freeze effect has been deactivated and the sensor can be easily withdrawn.
Never use excessive force to extract the temperature sensor.
Warning
DO NOT push the EXTRACTION button when the cryoprobe is not within the
target tissue, as skin burns could occur, and not before the freezing protocol
is completed, unless you want to shorten the procedure due to clinical
judgment.
Warning
Never detach the cryoprobe from the cryohandle if you are not clearly
instructed to unscrew or disengage it.
Caution
Verify cryoprobe S/N registration by double checking the serial number ON
the package and on the cryoprobe itself. Entering an incorrect cryoprobe S/N
registration will result in cryoprobe nullification.
Warning
Do not allow any liquid or humidity to enter the cryohandle. Always keep the
cover ON the cryohandle.
Warning
In case of the EXTRACTION process isn't available, Wait for passive THAW.
Warning
In case of frost on shaft, if possible, start EXTRACTION. If not, wait for passive
THAW. In both cases use skin protection techniques.