2.4.8
Mechanical handling of hose and cryohandle
Warning
Do not try to disengage a cryoprobe before instructed since the system is
under pressure and you can be hurt by the sharp cryoprobe ejected under
pressure.
Warning
The hose and the cryoprobe still contain liquid nitrogen that can be spilled
out if you disengage the cryoprobe before you get proper system
instructions to disengage the cryoprobe.
Warning
Never use excessive force to insert or remove a cryoprobe from the
cryohandle.
If moderate force is not sufficient, contact IceCure Medical for advice.
Caution
Do not pull the system by the cryohandle or hose.
The hose should be bent gently during procedure.
2.5
Emergencies and errors
2.5.1
Emergency Stop button
Caution
Only push the Emergency Stop button when there is NO other choice.
Whenever possible, use EXTRACTION to release the cryoprobe from the
tissue and use the standard shutdown procedure.
Excessive use of the Emergency Stop button may damage the system.
Warning
If a major electrical risk is detected during the cryotherapy procedure, PUSH
the Emergency Stop button immediately.