To avoid spillage and cold burns, do not transfer a Dewar containing liquid
nitrogen unless it is covered with its designated lid.
Warning
Do not use a liquid nitrogen Dewar if it is damaged.
You can tell that a Dewar is damaged if after filling it, frost appears on the outer
wall of the container. Return the Dewar to an IceCure Medical technician or an
authorized distributor for inspection.
Warning
At the beginning of the procedure the Dewar should be filled around 1 cm below
the neck of the bottle.
Do not overfill a Dewar with liquid nitrogen; it will reduce performance of the
system.
•
Take off the lid, and place the Dewar in its position into the system; bottom first.
Make sure the Dewar moves freely (right-to-left) when you move the Dewar-
handle (see pictures below).