Problem
Probable Cause
Action
Cryoprobe mechanical
damage
Bent or pinched cryoprobe.
1. Press the red emergency stop
button
2. Wait for 60 seconds.
3. If the cryoprobe is in the
patient’s body, wait for passive
thaw if necessary and remove
the cryoprobe gently.
•
Make sure that the cryoprobe
is removed intact.
4.
Make sure that the tip is
pointed in a safe direction and
then securely disengage the
cryoprobe.
5.
The system should not be
turned on until disengaging
the cryoprobe from the
cryohandle. The cryoprobe
should be screwed out of the
cryohandle.
6.
Close the handle plug on the
cryohandle, by screwing it in.
7.
Start the system, using the on-
off switch on the back side of
the system.
8.
If required, wait for the Dewar
to be lowered down and
perform a new procedure with
full Dewar and a new
cryoprobe while maintaining
sterility.
System provides “Used
Cryoprobe
” message.
System double use of the
cryoprobe.
Replace the cryoprobe.
“
Cryoprobe reading
failed” message
displayed on the screen.
RFID reader can’t find the tag.
Make sure cryoprobe is secured
and press ok to retry reading. If
the message reappears, change
the cryoprobe and call IceCure
Medical service.
FREEZE/EXTRACTION
icon does not initiate the
operation
Faulty icon or operator misuse.
Try pushing the icon again. It is
only activated if pushed for 1
second.
Check the user manual to ensure
that you can indeed use the icon in
the present screen.
If the problem persists contact
IceCure Medical service.
LED does not light up
Burned out LED, or it should not
be lighting up at given time.
Check user manual to verity that
the LED should light up.
If yes, contact IceCure Medical
service.