2.2.6
Installation and Setup
Warning
The cryoablation system should be operated in an adequately ventilated
room. Failure to do so may result in risk of suffocation due to increased levels
of nitrogen in the room.
Caution
After positioning the main chassis, lock the front roller brakes. Failure to do
so may result in damage to the system or to other equipment in the clinic
room.
Caution
The ProSense™ cryoablation system must be unpacked, installed, and tested
by an IceCure Medical authorized technician only.
Caution
There are no user-serviceable parts in the system. Refer all service issues to
IceCure Medical’s Customer Service Department.
Caution
Make sure the time is set according to local time zone before executing a
cryotherapy procedure.